Read: GUEST ON3DH
Subj: Re: S-Meter adjustment ?
Path: ON0AR<ON0AR<GB7CIP<GB7CRV<GB7FUR<GB7IOM<GB7SAN<GB7INV<GB7NOS
Sent: 050414/1457Z @:GB7NOS.#76.GBR.EU #:30985 [Golspie] FBB7.00i $:321898GM0CS
From: GM0CSZ@GB7NOS.#76.GBR.EU
To : IC706@WW
GM7HUD wrote:-
> 1. Turn power off, short out remote plug, hold down P.AMP/ATT and TUNER and
> turn the power on.
>
> 2. Press F2 to enter S-meter adj. (Use UP/DOWN to scroll the display)
>
> 3. Connect signal gen on 14.5151MHz, no signal and press MENU to set S0
> level.
>
> 4. Set siggen to 50uV output no modulation, push MENU to set S9 level.
>
> 5. Set siggen to 32mV output, no modulation, push MENU to set 60db level.
>
> 6. Push F1 to exit setup.
>
> Well that's what it says in the 703 manual. Try it at your own risk, E&OE.
=========================================
From the above it is my understanding that '60 dB' is meant to be S9 +
60dB
If that is the case then the voltage corresponding with S9 + 60dB is 1000
times the voltage applicable to S9 ; this equals 50 times 1000 is 50000
microvolts equals 50 mV. (not 32 mV)
60 dB or 6 Bell means a power ratio of 10E6 or a voltage ratio of sqrt10E6
equals 10E3 equals 1000.
This is NOT to be pedantic but to confirm my understanding or otherwise.
If I am wrong please let me know , I am open to learning.
73 - Frank , GM0CSZ @ GB7NOS.#76.GBR.EU